9
Digital multimeter specifications (continued)
Supplemental characteristics
DC voltage
Measurement method Sigma Delta A-to-D converter
Maximum input voltage 1000 V
dc
on all ranges
Input impedance 10 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF
Input protection 1000 V
rms
on all ranges
Response time
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
DC current
Measurement method Sigma Delta A-to-D converter
Maximum input current 10 mA to 3.0 A DC
1
Burden voltage and shunt resistance
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.05 Ω for 1 A range
• < 0.7 V, 0.05 Ω for 3 A range
Input protection Protected with 3.15 A/500 V, FF fuse
Response time
Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range
1. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a smaller current or offset current
measurement immediately after a high current measurement, ensure that the U3606B is cooled down.
AC voltage
Measurement method AC coupled true rms
Maximum input voltage 750 V
rms
/1200 V
peak
/3 x 10
7
V-Hz of product
Input impedance 1 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF
Input protection 750 V
rms
on all ranges
Crest factor
For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at
full scale.
Peak input 300% of range. Limited by maximum input.
Response time
Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested
input signal at the same range.
Overload ranging
Will select higher range if peak input overload is detected during auto range. Overload is
reported in manual ranging.