8
Duty cycle and pulse width specifications
Table 6. Duty cycle and pulse width resolution and accuracy
Function Range Resolution Accuracy of full scale
Duty cycle 100.000%
1
0.001% 0.3% + 0.2% per kHz
Pulse width
199.999 ms
2
1999.99 ms
2
0.001 ms
0.01 ms
Duty cycle/frequency
Duty cycle/frequency
1. The range is from {10 µs × frequency × 100%} to {[1 – (10 µs × frequency)] × 100%}. For example, a 1 kHz signal can be measured from
1% to 99%.
2. The positive or negative pulse width must be greater than 10 µs. The range of the pulse width is determined by the frequency of the signal.
Operating specifications
Table 7. Reading speed (typical)
1
Function Rate
Reading speed
2
(readings/second)
Reading speed
over USB
3
(readings/second)
Reading speed
over GPIB
4
(readings/second)
DC voltage
(10 V)
Slow (5½ digits)
Fast (4½ digits)
17
70
8
23
8
22
DC current
(1 A)
Slow (5½ digits)
Fast (4½ digits)
17
70
8
26
8
24
AC voltage
(10 V at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
17
70
8
23
8
22
AC current
(1 A at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
17
70
8
26
8
24
AC + DC voltage
(10 V at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
4
17
2.9
10
2.9
10
AC + DC current
(1 A at 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
4
17
2.9
10
2.9
10
Resistance
(100 kΩ)
Slow (5½ digits)
Fast (4½ digits)
17
70
8
22
8
22
Lo-Ω
(1 kΩ)
Slow (5½ digits)
Fast (4½ digits)
17
70
0.8
0.8
0.8
0.8
Capacitance
(10 µF)
Slow/Fast (3½ digits) 5 1.4 1.4
Diode
(1 V)
Slow/Fast (4½ digits) 70 26 23
Frequency
(voltage path at
10 V, 1 kHz)
Slow (5½ digits)
Fast (4½ digits)
9
9
8
8
8
8
Frequency
(current path at
1 A, 1 kHz)
Slow (5½ digit)
Fast (4½ digit)
9
9
8
8
8
8
1. Based on an average of 500 readings.
2. Reading rate of the A/D converter.
3. Number of measurements per second that can be read through USB using SCPI “READ?” command.
4. Number of measurements per second that can be read through GPIB using SCPI “READ?” command.
Digital multimeter specifications (continued)