26
General Purpose Differential Active Probes - N2792A/93A General-
purpose Differential Probes
N2792A 200-MHz, 20-V differential probe
N2793A 800-MHz, 15-V differential probe
N2792A N2793A
Bandwidth 200 MHz 800 MHz
Risetime 1.75 ns 437 ps
Attenuation ratio 10:1 10:1
CMRR 80 dB at 50/60 Hz
50 dB at 10 MHz
-60 dB at 50/60 Hz
-15 dB at 500 MHz
Max input voltage to ground ±60 V ±40 V
Max input voltage between two inputs ±20 V ±15 V
Characteristics for N2792A and N2793A differential probes
Model Number Description
1141A 200-MHz differential probe
1142A Probe control and power module for 1141A
1153A 200-MHz differential probe with AutoProbe interface
N2790A 100-MHz, 1.4 kV differential probe with AutoProbe interface
N2791A N2791A 25-MHz, 700-V differential probe
N2792A N2792A 200-MHz, 20-V differential probe
N2793A N2793A 800-MHz, 15-V differential probe
N2891A N2891A 70-MHz, 7,000-V differential probe
Ordering information for Agilent differential probes and power supply
The N2792A 200-MHz and N2793A
800-MHz differential probes provide the
superior general-purpose differential
signal measurements required for today’s
high-speed power measurements, vehicle
bus measurements, and digital system
designs.
The N2792A and N2793A probes offer
a 10:1 attenuation setting and high input
resistance and low input capacitance to
minimize circuit loading.
Both probes are compatible with any
oscilloscope with 50 Ω BNC input. The
probe can be powered by any USB port on a
scope or computer, or by a 9 V battery.