Chapter 8 191
Performance Tests
RF Generator Level Accuracy Performance Test 5
3. On the measuring receiver:
a. Set the measurement mode to RF Power.
b. Set the display to log.
4. On the Test Set:
a. Press
Preset.
b. Select the INSTRUMENT CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Select the CDMA GENERATOR screen.
e. Set the CW RF Path field to Bypass.
f. Select the RF GENERATOR screen.
g. Set the RF Gen Freq to 3 MHz.
h. Set the Amplitude to -10 dBm.
5. For each frequency in the Performance Test Record (PTR) do the
following:
a. Set the measuring receiver to measure frequency.
b. Set the Test Set level to
−10 dBm.
c. After the measuring receiver has acquired the signal, set the
measuring receiver to measure tuned RF level.
d. Measure the RF level at the levels shown in the PTR at the Test
Set's DUPLEX OUT port and compare the measured RF level to
the limits. If the measuring receiver displays the need to
recalibrate, press the CALIBRATE key and wait for calibration to
be completed.
e. Move the sensor module to the Test Set's RF IN/OUT port.
f. On the Test Set set the Output Port field to RF Out and repeat
the measurements for the levels shown in the PTR and compare
the measured RF level to the limits.
g. Move the sensor module back to the Test Set's DUPLEX OUT port
and set the Output Port to Dupl.