594 Chapter13
Softkey Descriptions
Softkey Descriptions
FM OFFSET FM Offset
DESCRIPTION Note that this function is available for 8593E, 8594E,
8594Q, 8595E, and 8596E spectrum analyzers
equipped with Option 102, AM/FM Speaker and TV
Sync Trigger, only.
Use
FM OFFSET to adjust the horizontal trace for
center-screen with no modulation on the carrier. This
function is useful for adjusting the carrier offset when
FM GAIN is set for a modulated signal with a small
frequency deviation. The
CAL FREQ self-calibration
routine calibrates an initial center-screen offset.
Use FM OFFSET to do a functional check of the
demodulation circuitry. The CAL OUT signal can be
used in place of an unmodulated carrier signal.
1. Connect the CAL OUT signal to the analyzer input
and make the following instrument settings.
PRESET
FREQUENCY..............................................300 MHz
SPAN............................................................0 Hz
BW................................................................5 MHz
AMPLITUDE ...............................................20 dBm
2. Press the following keys.
AUX CTRL
DEMOD
DEMOD ON OFF (ON)
DEMOD AM FM (FM)
CAL
MORE 1 of 3
MORE 2 of 3
SERVICE DIAG
MORE 1
MORE 2
MORE 3
MORE 4
FM OFFSET