Chapter 13 557
Softkey Descriptions
Softkey Descriptions
Understanding Analyzer Passcodes
Passcodes activate specific softkey functions and protect the correction
data from accidental erasure or modification. The two passcodes,
37 Hz and 2001 Hz, are explained below.
37 Hz Passcode
Use the 37 Hz passcode to perform the following functions.
Replace the current calibration-factors with the default calibration
data provided by
DEFAULT CAL DATA.
Check the accuracy of the 10 MHz frequency reference using
VERIFY TIMEBASE. Refer to the 10 MHz reference accuracy
verification test in the calibration guide for your instrument.
Bypass the CAL OUT set-up check within individual self-calibration
routines.
2001 Hz Passcode
Use the 2001 Hz service passcode to perform the following functions.
Set the start frequency, stop frequency, and step size of the flatness
correction points using
INIT FLT.
Edit the flatness-correction constants using
EDIT FLATNESS.
Edit the A12 amplitude control step-attenuator correction constants
using SET ATTN ERROR.
Set the correction factor for the 10 MHz frequency reference
(standard timebase).
Modify the displayed power units that appear when the instrument
is first turned on using STOR PWR ON UNITS.
Bypass the
CAL OUT signal check within individual self-calibration
routines.
Entering a Passcode
The passcode must be entered before pressing CAL. Press the following
keys to enter a passcode.
FREQUENCY
37, HZ or 2001, HZ
CAL
When SRVC appears in the lower-left corner of the display, the passcode
has been accepted.