c. Expand the outer slide mounts to their full length. Mount the
instrument onto the system cabinet by bolting the outer slide
mount to the inner-member bracket on the instrument, using
three
M5x0.8~12
flat-head screws per side.
Note
If any binding is encountered in the slides after mounting, it will be
necessary to slightly move the Unistrut nuts. While supporting the
instrument, loosen one of the Unistrut nuts. Adjust the slides slightly
until they operate freely. Retighten the Unistrut nut.
19. Mounting in a Non-HP System Rack Cabinet:
a. Using hole pattern “B,” see Figure 2-14, attach one adapter
bracket to the front of each outer slide mount, using two
M4x0.8~12
flat-head screws and nuts per bracket.
Figure 2-14. Slide Adapter for Non-HP System Rack Cabinets
2-18 Installation