E-Series power sensor specifications
The E-Series of power sensors have their calibration factors stored in EEPROM and operate over a wide dynamic range. They are designed
for use with the EPM Series of power meters and two classes of sensors are available:
CW power sensors (E4412A and E4413A).
Average power sensors (E9300 sensors).
E-Series CW power sensor specifications
Calibration factor (CF) and reflection coefficient (Rho)
Calibration factor and reflection coefficient
data are provided at 1 GHz increments on a
data sheet included with the power sensor.
This data is unique to each sensor. If you
have more than one sensor, match the serial
number on the data sheet with the serial
number on the power sensor you are using.
The CF corrects for the frequency response
of the sensor. The EPM power meter
automatically reads the CF data stored in the
sensor and uses it to make the corrections.
For power levels greater than 0 dBm, add
0.5%/dB to the calibration factor uncertainty
specification.
Reflection coefficient (Rho) relates to the
SWR according to the following formula: SWR
= 1 + Rho/1  Rho.
Maximum uncertainties of the CF data are
listed in table 5a, for the E4412A power
sensor, and table 5b for the E4413A power
sensor. The uncertainty analysis for the
calibration of the sensors was done in
accordance with the ISO/TAG4 Guide. The
uncertainty data reported on the calibration
certificate is the expanded uncertainty with
a 95% confidence level and a coverage
factor of 2.
Uncertainty*(%)
Reference
1.8
1.8
2.4
2.4
2.4
2.4
2.6
2.6
2.8
2.8
2.8
2.8
2.8
3.0
3.0
3.0
3.0
Frequency
50 MHz
100 MHz
1.0 GHz
2.0 GHz
4.0 GHz
6.0 GHz
8.0 GHz
10.0 GHz
11.0 GHz
12.0 GHz
14.0 GHz
16.0 GHz
17.0 GHz
18.0 GHz
20.0 GHz
24.0 GHz
26.0 GHz
28.0 GHz
Table 5b: E4413A calibration factor uncertainty at
1mW (0 dBm)
Table 5a: E4412A calibration factor uncertainty at
1mW (0 dBm)
Frequency
10 MHz
30 MHz
50 MHz
100 MHz
1.0 GHz
2.0 GHz
4.0 GHz
6.0 GHz
8.0 GHz
10.0 GHz
11.0 GHz
12.0 GHz
14.0 GHz
16.0 GHz
18.0 GHz
Uncertainty*(%)
1.8
1.8
Reference
1.8
1.8
2.4
2.4
2.4
2.4
2.4
2.4
2.4
2.4
2.6
2.6
Widest dynamic range: 100 pW to 100 mW (70 dBm to +20 dBm)
Table 4: E4410 Series max SWR specification
Model
E 4412A
E4413A
Maximum SWR
*10 MHz to <30 MHz: 1.22
30 MHz to <2 GHz: 1.15
2 GHz to <6 GHz: 1.17
6 GHz to <11 GHz:1.2
11 GHz to <18 GHz: 1.27
50 MHz to <100 MHz: 1.21
100 MHz to <8 GHz: 1.19
8 GHz to <18 GHz: 1.21
18 GHz to 26.5 GHz: 1.26
Maximum SWR
10 MHz  18 GHz
50 MHz  26.5 GHz
Maximum power
200 mW (+23 dBm)
200 mW (+23 dBm)
Connector type
Type-N (m)
APC-3.5 mm (m)
* Applies to sensors with serial prefix US 3848 or greater