Seitenansicht 69
Status Reporting Data Structures
Figure 6-2 brings together the different status reporting data structures
mentioned in this chapter and shows how they work together. To make it
possible for any of the Standard Event Status Register bits to generate a
summary bit, the bits must be enabled. These bits are enabled by using the
*ESE common command to set the corresponding bit in the Standard Event
Status Enable Register.
To generate a service request (SRQ) interrupt to an external controller, at
least one bit in the Status Byte Register must be enabled. These bits are
enabled by using the *SRE common command to set the corresponding bit in
the Service Request Enable Register. These enabled bits can then set RQS
and MSS (bit 6) in the Status Byte Register.
Status Reporting
Status Reporting Data Structures
6-5
Seitenansicht 69