Chapter 4 Command Reference
:STATus Subsystem
4-34 Programming Guide
4
COMMAND :STATus:OPERation:NTRansition <non-decimal numeric> | <NRf>
Sets or queries the negative transition filter for the Operation status
reporting structure.
RANGE The range of the <non-decimal numeric> or <NRf> parameter is
0 to 32,767.
QUERY RESPONSE Numeric data transferred as ASCII bytes in <NR1> format.
COMMENTS The parameter and the query response value, when rounded to an
integer value and expressed in base 2 (binary), represent the bit values
of the negative transition filter.
The value of unused bits is zero when queried and is ignored when set.
A TRUE bit in the negative transition filter specifies that a negative
(TRUE to FALSE) transition of the corresponding bit in the Operation
Condition Status Register generates the corresponding event in the
Operation Event Status Register.
At power-on and STAT:PRES, the negative transition filter is preset
such that each bit is a 0 (FALSE).
This value is unaffected by *RST, *CLS, and *SAV/*RCL.
RELATED
FRONT-PANEL
KEYS
None
COMMAND :STATus:OPERation:PTRansition . . . <non-decimal numeric> | <NRf>
Sets or queries the positive transition filter for the Operation status
reporting structure.
RANGE The range of the <non-decimal numeric> or <NRf> parameters is
0 to 32,767.
QUERY RESPONSE Numeric data transferred as ASCII bytes in <NR1> format.
Seitenansicht 129