Agilent Technologies 53150A Bedienerhandbuch Seite 105

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 174
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 104
Chapter 4 Command Reference
:INPut Subsystem
Programming Guide 4-9
4
COMMAND :INITiate[:IMMediate]
This event command causes the instrument to initiate either a single
measurement or a block of measurements.
COMMENT This command is an overlapped command (see IEEE 488.2, Section 12).
Beginning a measurement or block of measurements with an
:INITiate[:IMMediate] sets the Pending Operation Flag to true.
Completing the measurement or block of measurements (normally or
by aborting) sets Pending Operation Flag to false.
RELATED
FRONT-PANEL
KEYS
None
:INPut Subsystem
This subsystem controls the characteristics of the Counter’s input ports.
COMMAND :INPut:FILTer:[:LPASs][:STATe] . . . <Boolean>
Sets or queries the state of the Channel 1 low-pass filter.
QUERY RESPONSE Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
COMMENT *RST: OFF
RELATED
FRONT-PANEL
KEYS
Shift + Menu (CH1 LPF > OFF | ON)
Seitenansicht 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 173 174

Kommentare zu diesen Handbüchern

Keine Kommentare