2 General Power Meter Functions
34 N1913/1914A EPM Series Power Meters User’s Guide
Setting the Resolution
The resolution of each of the power meter’s numeric type windows can be
set to four different levels (1, 2, 3 or 4).
These four levels represent:
1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or
dB.
1, 2, 3 or 4 significant digits respectively if the measurement suffix is
W or %.
The default value is 0.01 dB (3 digits).
To set the resolution on the currently selected window:
1 Press . The current setting of the resolution is highlighted under
the
softkey.
2 To change this setting, press repeatedly until the required
resolution setting is highlighted.
Resolution
Resolution