6 Using N8480 Series Power Sensors
160 N1913/1914A EPM Series Power Meters User’s Guide
Selecting a Sensor Calibration Table
You can select a calibration factor table from the key menu
followed by and .
The State column indicates if any calibration factor tables are currently
selected. The Sensor Tbls screen is shown in Figure 6- 8.
Procedure
Select sensor calibration table as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and
press to highlight .
Figure 6-8 Sensor table selected
Tab les
Sensor Cal Table
NOTE
You can also view which sensor table is being used by pressing the , and
use the and keys to highlight the CF Table setting field and press
to display the table.
Offset
Tab les
Sensor Cal Tables
Tab le
On