Resistance Measurement Configuration
The meter can make 2-wire or 4-wire ohms measurements. The meter
can compensate for voltages in the resistive circuit being measured using
offset compensation. You can also choose to either limit the power applied
or limit the open circuit voltage applied during a resistance
measurement.
The test current sourced by the meter depends upon the range and
conditions as shown below:
Range Test current
2-wire or 4-wire
measurements
4-wire measurements
Low Power Voltage Limited
1
10 mA 10 mA Not Available
10
10 mA 10 mA 1 mA
100
10 mA 1 mA
100 µA
1 k
1 mA
100 µA
Not Available
10 k 100 µA10 µA
Not Available
100 k*10 µA5 µA
Not Available
1 M*5 µA5 µA
Not Available
* Offset Compensation is not applied to measurements on these ranges