Service GuidePublication Number 54622-97037December 2002For Safety Information, Warranties, and Regulatory information, see the pages at the end of th
1-4General Information• Agilent IntuiLink Data Capture (for 54622A, 22D, or 24A)IntuiLink Data Capture is a standalone program for downloading wavefor
6-10Replacing AssembliesTo remove the system boardTo remove the system board1 Remove the floppy drive.2 Using the T10 TORX driver, remove the six scre
6-11Replacing AssembliesTo remove the system board7Lift the back of the board to clear the main deck and then pull the board straight out.Figure 6-8Re
6-12Replacing AssembliesTo remove the power supplyTo remove the power supply1 Remove the fan.2 Disconnect the ground wire (green wire with the yellow
6-13Replacing AssembliesTo remove the power supply5Use a flat-blade screwdriver to gently unhook the latch that holds the white power-actuator shaft t
6-14Replacing AssembliesTo remove the power supply6Using the T20 TORX driver, remove the screw holding the power supply board to the deck. Figure 6-11
6-15Replacing AssembliesTo remove the keyboard assemblyTo remove the keyboard assembly1 Remove the front panel. 2 Remove all of the knobs by pulling t
6-16Replacing AssembliesTo remove the handleTo remove the handle• Rotate the handle downward until it just passes the last detente position; this is a
7Replaceable Partsservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
7-2Replaceable PartsThis chapter describes how to order replaceable assemblies and parts for the Agilent 54621A/22A/24A Oscilloscope and Agilent 54621
7-3Replaceable PartsTo order a replacement partTo order a replacement part• To order a part in the material list, quote the Agilent Technologies part
1-5General InformationFigure 1-1Package contents for 54620-Series Oscilloscopesss110074C ProbesPower cordAccessories pouch and front-panel cover**5462
7-4Replaceable PartsTo order a replacement partFigure 7-1Exploded View of the 54621A/22A/24A/21D/22D Oscilloscopeservice.book Page 4 Wednesday, Dece
7-5Replaceable PartsTo order a replacement partTable 7-1Replaceable Parts Reference DesignatorAgilent Part NumberQty DescriptionA1 54621-69001 1 Refur
7-6Replaceable PartsTo order a replacement partMP1 54801-47401 See Descr Knob - 12mm Flint Gray (54621A/22A – 6 ea)(54621D/22D/24A – 8 ea)MP2 54801-
7-7Replaceable PartsTo order a replacement partMP25 54622-63403 See Descr Attenuator lid(54621A/21D/22A/22D - 1 each)(54624A - 2 each)W1 54630-61601 1
7-8service.book Page 8 Wednesday, December 18, 2002 8:35 AM
DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014Manufacturer’s Name:Manufacturer’s Address:Agilent Technologies, Inc.Ga
Product RegulationsEMCIEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998CISPR 11:1990 / EN 55011:1991IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995IEC
Agilent Technologies Inc.P.O. Box 21971900 Garden of the Gods RoadColorado Springs, CO 80901-2197, U.S.A.Safety NoticesThis apparatus has been designe
Notices© Agilent Technologies, Inc. 2000-2002No part of this manual may be reproduced in any form or by any means (including electronic storage and re
1-6General InformationTo inspect options and accessories❏ Verify that you received the options and accessories you ordered and that none were damaged.
1-7General InformationTable 1-2Accessories availableModel Description01650-61607 16:16 logic cable and terminator (for use with 54621D/22D)54620-68701
1-8General InformationTable 1-3. Power Cords Plug Type Cable Part Number Plug Type Cable Part NumberOpt 900 (U.K.) 8120-1703 Opt 918 (Japan) 8120-475
1-9General InformationAcquisition: Analog ChannelsPerformance Characteristics* Denotes Warranted Specifications, all others are typical. Specificatio
1-10General InformationVertical System: Analog ChannelsVertical System: Analog ChannelsAnalog channels 54621A/21D, 54622A/22D: Ch1 and 2 simultaneous
1-11General InformationVertical System: Analog Channels (continued)Vertical System: Analog Channels (continued)ESD Tolerance ±2 kVNoise Peak-to-Peak 2
1-12General InformationHorizontalHorizontalRange 5 ns/div to 50 s/divResolution 25 psVernier 1-2-5 increments when off, 25 minor increments between ma
1-13General InformationTrigger SystemTrigger SystemSources: 54621A/22A: Ch 1, 2, line, ext54621D/22D: Ch 1, 2, line, ext, D15 - D054624A: Ch 1, 2, 3,
iiThe Oscilloscope At a GlanceChoose from a variety of oscilloscopes for capturing long, non-repeating signalswith 200 MSa/s sample rate and 2 MBytes
1-14General InformationAnalog Channel TriggeringAnalog Channel Triggering Range (Internal) ±6 div from center screenSensitivity* Greater of 0.35 div o
1-15General InformationDisplay SystemDisplay SystemDisplay 7-inch raster monochrome CRTThroughput of Analog Channels 25 million gray scale vectors/sec
1-16General InformationFFTFFTPoints Fixed at 2048 pointsSource of FFT Analog channels 1 or 2 (or 3 or 4 for 54624A), 1+2, 1-2, 1*2Window Rectangular,
1-17General InformationGeneral CharacteristicsGeneral CharacteristicsPhysical:SizeWeight32.26 cm wide x 17.27 cm high x 31.75 cm deep (without handle)
1-18service.book Page 18 Wednesday, December 18, 2002 8:35 AM
2Preparing the Oscilloscope for Useservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
2-2Preparing the Oscilloscope for UseTo prepare your oscilloscope for use, you need to do the following tasks. After y ou have completed them, you wil
2-3Setting up the OscilloscopeAfter you have done a few basic tasks, you will connect probes to the oscilloscope. The number of probes, and the type o
2-4Preparing the Oscilloscope for UseTo adjust the handleTo adjust the handle1 Grasp the handle pivot points on each side of the instrument and pull t
2-5Preparing the Oscilloscope for UseTo power-on the oscilloscopeTo power-on the oscilloscope1 Connect the power cord to the rear of the oscilloscope,
iiiIn This BookThis book provides the service information for the Agilent 54621A/22A/24A Oscilloscope and the Agilent 54621D/22D Mixed-Signal Oscillos
2-6Preparing the Oscilloscope for UseTo adjust the display intensityTo adjust the display intensityThe Intensity control is at the lower left corner o
2-7Preparing the Oscilloscope for UseTo connect the oscilloscope probesTo connect the oscilloscope probes1 Connect the Agilent 10074C 1.5-meter, 10:1
2-8Preparing the Oscilloscope for UseTo use the digital probes (mixed-signal oscilloscope only)To use the digital probes (mixed-signal oscilloscope on
2-9Preparing the Oscilloscope for UseTo use the digital probes (mixed-signal oscilloscope only)3 Connect a grabber to one of the probe leads. Be sure
2-10Preparing the Oscilloscope for UseTo use the digital probes (mixed-signal oscilloscope only)5 For high-speed signals, connect a ground lead to the
2-11Preparing the Oscilloscope for UseTo use the digital probes (mixed-signal oscilloscope only)7 Repeat steps 3 through 6 until you have connected al
2-12Preparing the Oscilloscope for UseTo connect a printerTo connect a printerThe oscilloscope connects to a parallel printer through the Parallel out
2-13Preparing the Oscilloscope for UseTo verify basic oscilloscope operationTo verify basic oscilloscope operation1 Connect an oscilloscope probe to c
2-14Getting started using the oscilloscope interfaceWhen the oscilloscope is first turned on, it performs a self-test, then momentarily shows a startu
2-15Preparing the Oscilloscope for UseTo verify basic oscilloscope operation•Press the Getting Started softkey to view the symbols used in the oscillo
ivservice.book Page iv Wednesday, December 18, 2002 8:35 AM
2-16Using Quick HelpThe oscilloscope has a Quick Help system that provides user help for each front-panel key and softkey on the oscilloscope. To view
2-17Preparing the Oscilloscope for UseSelecting a language for Quick Help after you have been operating the oscilloscopeSelecting a language for Quick
2-18Preparing the Oscilloscope for UseLoading a language from floppy diskLoading a language from floppy diskLanguage files can be downloaded from www.
2-19Cleaning the oscilloscope1 Disconnect power from the instrument.CAUTION Avoid Damage to Sensitive Electronic Components!Do not use too much liqu
2-20service.book Page 20 Wednesday, December 18, 2002 8:35 AM
3Testing Performanceservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
3-2Testing PerformanceThis chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the p
3-3Testing PerformanceList of Test EquipmentList of Test EquipmentBelow is a list of test and equipment and accessories required to perform the perfor
3-4Testing PerformanceTo construct the test connectorTo construct the test connectorThe Agilent 54621D/22D Mixed-Signal Oscilloscope has digital chann
3-5Testing PerformanceTo test the 54621D/22D Oscilloscope digital channelsTo test the 54621D/22D Oscilloscope digital channelsThe acquisition system t
ContentsContents-11 General InformationTo inspect package contents 1-3To inspect options and accessories 1-6Performance Characteristics 1-92 Preparing
3-6Testing PerformanceTo verify digital channel threshold accuracyTo verify digital channel threshold accuracyThis test verifies the digital channel t
3-7Testing PerformanceTo verify digital channel threshold accuracy1Turn on the test equipment and the oscilloscope. Let them warm up for 30 minutes be
3-8Testing PerformanceTo verify digital channel threshold accuracyb Method 2 — Using the Test FixtureUse the Agilent 01660-63801 Performance Verificat
3-9Testing PerformanceTo verify digital channel threshold accuracy7Press the oscilloscope User softkey, then turn the Entry knob () on the front panel
3-10Testing PerformanceTo verify voltage measurement accuracyTo verify voltage measurement accuracyThis test verifies the voltage measurement accuracy
3-11Testing PerformanceTo verify voltage measurement accuracyb Set the Volts/Div setting to the value in the first line in Table 3-5. Table 3-5Setting
3-12Testing PerformanceTo verify voltage measurement accuracy3Use the BNC tee and cables to connect the oscilloscope calibrator /power supply to both
3-13Testing PerformanceTo verify bandwidthTo verify bandwidthThis test verifies bandwidth. In this test you will use an oscilloscope calibrator with a
3-14Testing PerformanceTo verify bandwidth1Connect the oscilloscope calibrator output through a 50Ω feedthrough to the oscilloscope channel 1 input.2
3-15Testing PerformanceTo verify horizontal Dt and 1/Dt accuracyTo verify horizontal ∆t and 1/∆t accuracyThis test verifies the horizontal ∆t and 1/∆t
ContentsContents-24 Calibrating and AdjustingTo adjust the power supply 4-4To perform User Cal 4-7To adjust the oscilloscope display 4-85 Troubleshoot
3-16Testing PerformanceTo verify horizontal Dt and 1/Dt accuracy4Change the calibrator to 1-µs markers. Change the time base to 200 ns/div. Adjust the
3-17Testing PerformanceTo verify trigger sensitivityTo verify trigger sensitivityThis test verifies the trigger sensitivity. In this test, you will ap
3-18Testing PerformanceTo verify trigger sensitivityTest Internal Trigger Sensitivity1Press the Save/Recall key, then press the Default Setup softkey.
3-19Testing PerformanceTo verify trigger sensitivityTest External Trigger SensitivityVerify the external trigger sensitivity at these settings:100 MHz
3-20Agilent 54622A/22D/24A Performance Test RecordAgilent 54622A/54622D/54624ASerial No. ______________________________________ Test by ______________
3-21Agilent 54621A/21D Performance Test RecordAgilent 54621A/54621DSerial No. ______________________________________ Test by _________________________
3-22service.book Page 22 Wednesday, December 18, 2002 8:35 AM
4Calibrating and Adjustingservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
4-2Calibrating and AdjustingThis chapter explains how to adjust the oscilloscope for optimum operating performance. You should perform the hardware ad
4-3Calibrating and AdjustingWARNING HAZARDOUS VOLTAGES !Read the safety summary at the back of this book before proceeding.Maintenance is performed w
1General Information service.book Page 1 Wednesday, December 18, 2002 8:35 AM
4-4Calibrating and AdjustingTo adjust the power supplyTo adjust the power supplyThe oscilloscope power supply has both a +3.3 V adjustment, and a +5.1
4-5Calibrating and AdjustingTo adjust the power supply2Locate the power supply voltages at E204, E205, E207, and E208 on the system board. The power s
4-6Calibrating and AdjustingTo adjust the power supply4If the +5.1 V measurement is out of tolerance, adjust the +5.1 V / -5.2 V balance adjustment on
4-7Calibrating and AdjustingTo perform User CalTo perform User CalPerform self-calibration:• every 6 months or after 1000 hours of operation.• if the
4-8Calibrating and AdjustingTo adjust the oscilloscope displayTo adjust the oscilloscope displayIn this procedure, you will make adjustments using two
4-9Calibrating and AdjustingTo adjust the oscilloscope display5Adjust H.HOLD (horizontal hold) to center the display horizontally.6 Adjust FOCUS to ac
4-10service.book Page 10 Wednesday, December 18, 2002 8:35 AM
5Troubleshootingservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
5-2TroubleshootingThis chapter begins with suggestions for solving general problems that you may encounter with the oscilloscope. It tells you what to
5-3Solving General Problems with the OscilloscopesThis section describes how to solve general problems that you may encounter while using the 54621A/2
1-2General InformationThis chapter lists general information for the Agilent 54620-series Oscilloscopes. It also includes performance characteristics
5-4TroubleshootingIf the trace display is unusual or unexpectedIf the trace display is unusual or unexpected❏ Check that the Horizontal time/division
5-5TroubleshootingIf you cannot see a channelIf you cannot see a channel❏ Check that the oscilloscope probe cable is securely connected to the input c
5-6Troubleshooting the OscilloscopeThe service policy for the 54621A/22A/24A Oscilloscopes and the 54621D/22D Mixed-Signal Oscilloscopes is instrument
5-7TroubleshootingTo construct your own dummy loadTo construct your own dummy load1 Obtain a connector that is compatible with the connector J3 on the
5-8TroubleshootingTo check out the oscilloscopeTo check out the oscilloscope1 Is an interface module connected to the rear panel of the oscilloscope?
5-9TroubleshootingTo check out the oscilloscope5Disconnect the display ribbon cable from the system board. Then check the signals on the system board
5-10TroubleshootingTo check out the oscilloscope8Is the fan running? The low voltage power supply has a thermal cut-out circuit. If the fan is defecti
5-11TroubleshootingTo check the Low Voltage Power SupplyTo check the Low Voltage Power Supply 1 Disconnect the power cord from the oscilloscope. Then
5-12TroubleshootingTo check the Low Voltage Power SupplyTable 5-4 Low Voltage Power Supply Tolerances 4If the +5.1 V measurement is out of tolerance,
5-13TroubleshootingTo run the internal self-testsTo run the internal self-testsSelf Test performs a series of internal procedures to verify that the o
1-3General InformationTo inspect package contents❏ Inspect the shipping container for damage.If your shipping container appears to be damaged, keep th
5-14service.book Page 14 Wednesday, December 18, 2002 8:35 AM
6Replacing Assembliesservice.book Page 1 Wednesday, December 18, 2002 8:35 AM
6-2Replacing AssembliesThis chapter describes how to remove assemblies from the oscilloscope. After you have removed an assembly, to install the repla
6-3Replacing AssembliesWARNING HAZARDOUS VOLTAGES !Read the safety summary at the back of this book before proceeding.Maintenance is performed with p
6-4Replacing AssembliesTo remove the cabinetTo remove the cabinet1 Turn off the oscilloscope and disconnect the power cable. 2 Rotate the handle up as
6-5Replacing AssembliesTo remove the fanTo remove the fan1 Disconnect the fan cable from the power supply board. 2 Using the T20 TORX driver, remove t
6-6Replacing AssembliesTo remove the floppy driveTo remove the floppy drive1 Using the T8 TORX driver, loosen the two screws that hold the drive to th
6-7Replacing AssembliesTo remove the front panelTo remove the front panel1 Remove the intensity knob and shaft by grasping the knob with one hand and
6-8Replacing AssembliesTo remove the displayTo remove the display1 Remove the front panel from the oscilloscope.2 Disconnect the ribbon cable from the
6-9Replacing AssembliesTo remove the display4Use a flat-blade screwdriver to gently unhook the latch that holds the white power-actuator shaft to the