Installation Note 08720-90318 11
Verify the Frequency Reference Operation
1. Connect the equipment as shown in Figure 7.
Figure 7 Frequency Reference Operation Test
2. Press .
3. Connect the BNC to BNC jumper (supplied with the upgrade kit) between the EXT REF
and the 10 MHz High Stability Reference connectors. (Refer to Figure 8.)
Figure 8 Location of the BNC to BNC Jumper and Adjustments
4. Use a flat-head screwdriver to remove the screw that secures the high stability frequency
adjustment as shown in Figure 8. Insert a narrow screwdriver and adjust the high stability
frequency reference potentiometer for a reading of 50 MHz ± 5 Hz.
System Menu
CW FREQ
50
M/µ
Seitenansicht 10