Using N8480 Series Power Sensors 8
N1911A/1912A P-Series Power Meters User’s Guide 209
Selecting a Sensor Calibration Table
You can select a calibration factor table from the key menu
followed by and .
The State column indicates if any calibration factor tables are currently
selected. The Sensor Tbls screen is shown in Figure 8- 141.
Procedure
Select sensor calibration table as follows:
1 Press , , .
2 Use the and keys to highlight one of the 20 table titles and press
to highlight .
Figure 8-141Sensor table selected
Tab les
Sensor Cal Table
NOTE
You can also view which sensor table is being used by pressing the ,
and
use the and keys to highlight the CF Table setting field and press
to display the table.
Offset
Tab les
Sensor Cal Tables
Tab le
On