Agilent Technologies 1912A Betriebsanweisung Seite 135

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 297
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 134
General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 109
Offset and relative settings are ignored during calibration.
Manual Calibration
To manually calibrate the power meter and sensor combination:
1 Press and the channel softkey to start the calibration
routine.
The Calibrating pop- up is then displayed.
Tip You can reduce the steps required to carry out the manual zero and
calibration procedure as follow:
Press and . (For dual channel meters, press
,
or as required).
Cal
NOTE
For N1912A, you calibrate each channel independently by pressing the softkey by
choosing or .
Cal
Cal A
Cal B
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B
Seitenansicht 134
1 2 ... 130 131 132 133 134 135 136 137 138 139 140 ... 296 297

Kommentare zu diesen Handbüchern

Keine Kommentare