Seitenansicht 14
Agilent 44476A
Switching signals from
DC to 18 GHz
Triple 1-to-2 microwave
multiplexers
The 44476A includes three
independent SPDT 50 coaxial
relays with excellent electrical
performance from DC to 18 GHz.
For general-purpose microwave
switching applications, the
module can be used to switch
separate signal sources for a
multi-band receiver/transmitter
testing application. The 3 mm
SMA connector on the module
edge simplifies wiring.
Specifications
Input Characteristics
Frequency Range: DC to 18 GHz
Characteristic
Impedance: 50
Input Power Rating: 1 W average
(Also less than ± 7 VDC) 100 W peak
Repeatability
(after 10
6
operation): 0.03 dB
Connector: SMA
AC Isolation / Performance
Isolation: DC-18 GHz >90 dB
Insertion Loss: DC-2 GHz <0.25 dB
DC-18 GHz <0.50 dB
VSWR DC-2 GHz <1.15 dB
(3 mm SMA):
DC-12.4 GHz <1.25 dB
DC-18.0 GHz <1.40 dB
Agilent 44476B
Supporting varieties of
microwave coaxial relays
Two set mounting holes for
coaxial relays
The 44476B brings multi-port
50/75 coaxial switching
flexibility to your test system.
It has two set mounting panels,
so any two Agilent 876XX
coaxial switches can be mounted.
The coaxial switches come in
three-, four-, and five-port
configurations. This flexibility
allows you to use the different
switches for a variety of
applications, constructing
transfer switches, switch
matrices and more. Using the
876XX in conjunction with the
44476B allows you to extend
your automated three-port
switching to 26.5 GHz. Coaxial
relays must be ordered separately
when using with this module.
The coaxial switches that can be
used are listed below. Option 011
designates the switches for a
coil voltage of 5 VDC.
Agilent Technologies
Coaxial Switch Port Frequency
8762A/Option 011 3 DC to 4 GHz
8762B/Option 011 3 DC to 18 GHz
8762C/Option 011 3 DC to 26.5 GHz
8762F/Option 011 3 DC to 4 GHz
8763B/Option 011 4 DC to 18 GHz
8763C/Option 011 4 DC to 26.5 GHz
8764B/Option 011 5 DC to 18 GHz
8764C/Option 011 5 DC to 26.5 GHz
Note: For details of Agilent 876XX specifications,
please refer to publication number 5968-4314.
Microwave Multiplexer Module
Microwave Relay Driver Module
Form C
Drive Relays
Drive
Jumpers
Relay
K901
COM
NO
Channel 00
+5
1
W1
2
2
1
2 NO
C C
1 NC
1 = PORT 1
2 = PORT 2
NO = NORMALLY OPEN
NC = NORMALLY CLOSED
C = COMMON
* NOTE: Terminal designations inside the dashed boxes refer to the HP8762B only.
Module panel designators outside the dashed boxes are those silkscreenedon the
HP44476A panel. Any reference in this manual to a port refers to the module panel designators.
3
ORG
HP8762B
50Ω
P
I
V
O
T
A
R
M
A
T
U
R
E
YEL
ORG
NC
MODULE*
PAN EL
DESIGNATORS
Relay
K902
COM
NO
Channel 01
+5
1
W2
2
2
1
2 NO
C C
1 NC3
ORG
HP8762B
50Ω
P
I
V
O
T
A
R
M
A
T
U
R
E
YEL
ORG
NC
Relay
K903
COM
NO
Channel 02
+5
1
W3
2
2
C
C
C
1
2 NO
C C
1 NC3
ORG
HP8762B
50Ω
50Ω
50Ω
50Ω
+5V
+5 TO DRIVE JUMPERS (YEL)
L901
C906
P
I
V
O
T
A
R
M
A
T
U
R
E
YEL
ORG
NC
Form C
Drive Relays
Drive
Jumpers
Relay
K901
COM
NO
Channel 0
0
+5
1
W1
2
3
ORG
USER
INSTALLED
MICROWAVE
SWITCH
YEL
ORG
NC
Relay
K902
COM
NO
Channel 01
+5
1
W1
2
3
ORG
USER
INSTALLED
MICROWAVE
SWITCH
YEL
ORG
NC
Relay
K903
COM
NO
+5
+5V
+5 TO DRIVE JUMPERS (YEL)
L901
C906
1
2
3
CHANNEL 02
AVAILABLE
BUT NOT
USED
NC