Limit lines of mask template shown in Figure 7
Mask Template for Inside failure condition operating hint
When selecting the failure condition to be inside, an extra requirement is
placed on the mask template. The mask template can only do testing with both
maximum and minimum limits present. Therefore, at a certain time bucket, if
one of the two limit lines is set to be "Don’t Care," this time bucket is not tested.
When the signal fails the test, the action message for Fail Min is displayed.
Figure 8
Operating the Test Automation Module
To set up the mask template testing for "Fail when In" testing
43