Agilent B2900 SCPI Command Reference, Edition 2 5- 11
Error Messages
Query Error
Query Error
If the output queue control of Agilent B2900 detects one of following problems, a
-4XX error occurs.
An attempt was made to read data from the output queue when no output data is
present or pending.
Data in the output queue has been lost.
Error -400 Query error
Generic query error for B2900 that cannot be determined more specifically.
Error -410 Query INTERRUPTED
A condition causing an INTERRUPTED query error occurred; for example, a query
followed by DAB or GET before a response was completely sent.
Error -420 Query UNTERMINATED
A condition causing an UNTERMINATED query error occurred; for example,
B2900 was addressed to talk and an incomplete program message was received.
Error -430 Query DEADLOCKED
A condition causing a DEADLOCKED query error occurred; for example, both
input buffer and output buffer are full and B2900 cannot continue.
Error -440 Query UNTERMINATED after indefinite response
A query was received in the same program message after a query requesting an
indefinite length response was executed.
Seitenansicht 296