Agilent Technologies B2902 Wartungshandbuch Seite 146

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 326
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 145
4- 24 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:DISPlay:ENABle
digits=6 selects 5½ digit resolution.
digits=7 selects 6½ digit resolution.
Query response digits <newline>
digits returns the present setting. If a parameter is specified, digits returns the value
assigned to DEF, MIN, or MAX. Response data type is NR1.
Example :DISP:DIG 4
:DISP:DIG? MAX
:DISPlay:ENABle
Enables or disables the front panel display under remote operation. This command
setting is not changed by power off or the *RST command.
Regardless of this command setting, the front panel keys and the rotary knob are
disabled during remote operation. However, only the Local key is effective for
returning the instrument to local operation.
Syntax :DISPlay:ENABle mode
:DISPlay:ENABle?
Parameter mode 0|OFF|1|ON. Parameter data type is boolean.
mode=1 or ON enables the front panel display.
mode=0 or OFF disables the front panel display.
Query response mode <newline>
mode is 0 or 1, and indicates that front panel display is off or on, respectively.
Response data type is NR1.
Example :DISP:ENAB OFF
:DISP:ENAB?
:DISPlay:VIEW
Sets the display mode, single 1, single 2, dual, graph, or roll.
Syntax :DISPlay:VIEW mode
Seitenansicht 145
1 2 ... 141 142 143 144 145 146 147 148 149 150 151 ... 325 326

Kommentare zu diesen Handbüchern

Keine Kommentare