Keysight B1500A User’s Guide, Ed ition 14 4- 11
Specifications
Hardware
ASU
ASU: Atto Sense and Switch Unit
Maximum voltage, AUX path
100 V (AUX input to AUX common)
100 V (AUX input to circuit common)
42 V (AUX common to circuit common)
Maximum current, AUX path
0.5 A (AUX input to Force output)
Band width (at -3 dB), AUX port
30 MHz, supplemental characteristics
NOTE
Connect ASU to dedicated SMU
The specifications of the ASU are guaranteed while it is connected to the exclusive
SMU which is connected together when the calibration is performed. Confirm the
serial number of the ASU, and connect it properly.
NOTE
Connection path control
Disable source output of the instruments connected to the ASU inputs before
controlling the ASU connection path.