Measurement supplemental characteristics


Over range :
102 % of measurement range for 1.5 A and 3 A ranges.
106 % of measurement range other than 1.5 A and 3 A ranges.
Voltage measurement range change overshoot:

Current measurement range change overshoot:
< 250 mV/R load, 20 MHz bandwidth
Derating accuracy for measurement speed less than 1 PLC :
Add % of range using the following table for measurement with PLC < 1.
Derating accuracy with PLC setting< 1 PLC
Voltage range Current range
0.2 V 2 V to 200 V 10 nA 100 nA 1 μA to 100 mA 1 A to 3 A
0.1 PLC 0.01 % 0.01 % 0.1 % 0.01 % 0.01 % 0.01 %
0.01 PLC 0.05 % 0.02 % 1 % 0.1 % 0.05 % 0.02 %
0.001 PLC 0.5 % 0.2 % 5 % 1 % 0.5 % 0.2 %

Triggering
Digital I/O trigger in to trigger out
1

Digital I/O trigger in to source change
1

LXI trigger in to source change
1

LXI trigger in to measurement
1

Internal event to external LXI trigger output
1

Maximum Unknown
LAN[0-7] Event Receiving Latency
1
: Unknown
LAN[0-7] Event Sending Latency
1
: Unknown

1. Supplemental characteristics
Timer
 
Timer accuracy: ±50 ppm
Arm/trigger delay: 0 to 100000 s
 

Arm/trigger event: 1 to 100000
Time stamp: TIMER value automatically saved when each measurement is
triggered.