Program and interface capabilities, continued
LXI
LAN protocols and adhere to LXI requirements such as a built-in Web control server and
IVI-COM driver.
Ethernet: 10/100Base-T
USB2.0: USB-TMC488 protocol (Rear x 1)
GPIB: IEEE-488.2 compliant
USB 2.0 high-speed mass storage (MSC) class device (Front x 1)
Digital I/O interface
Connector: 25-pin female D.
Input/output pins: 14 open drain I/O bits
Absolute max input voltage: 5.25 V
Absolute min input voltage: –0.25 V
Max logic low input voltage: 0.8 V
Min logic high input voltage: 2.0 V
Max source current: 1 mA, Vout = 0 V
Max sink current: 50 mA, Vout = 5 V
5 V power supply pin: Limited to 500 mA, solid state fuse protected.
Safety interlock pin: One active high pin and one active low pin.
Activation of both pin enables output voltage > 42 V.
Maximum number of simultaneously triggered units (using Digital I/O)
1
: 8
1. Supplemental characteristic