1-77
Making Measurements
Using Limit Lines to Test a Device
2. To establish the start frequency and limits for a sloping limit line that tests the low side
of the filter, press:
3. To terminate the lines and create a sloping limit line, press:
4. To establish the start frequency and limits for a sloping limit line that tests the high
side of the filter, press:
5. To terminate the lines and create a sloping limit line, press:
You could use this type of limit to test the shape factor of a filter.
Figure 1-61 Sloping Limit Lines
ADD
STIMULUS VALUE
123 M/µ
UPPER LIMIT
65 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
125 M/µ
UPPER LIMIT
26 x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN
ADD
STIMULUS VALUE
144 M/µ
UPPER LIMIT
26 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
146 M/µ
UPPER LIMIT
65 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN