Agilent Technologies 24A Betriebsanweisung Seite 328

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 453
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 327
6-40
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
Calibrating for Noninsertable Devices
A test device having the same sex connector on both the input and output cannot be
connected directly into a transmission test configuration. Therefore, the device is
considered to be noninsertable, and one of the following calibration methods must be
performed:
adapter removal (ES Analyzers Only)
matched adapters
modify the cal kit thru definition
Figure 6-12 Noninsertable Device
Adapter Removal (ES Analyzers Only)
The adapter removal technique provides a means to accurately measure noninsertable
devices. The following adapters are needed:
Adapter A1, which mates with port 1 of the device, must be installed on test set port 1.
Adapter A2, which mates with port 2 of the device, must be installed on test set port 2.
Adapter A3 must match the connectors on the test device. The effects of this adapter
will be completely removed with this calibration technique.
NOTE Adapters A1 and A2 become part of the test setup to allow connection to the
DUT.
Adapter A3 is used during the calibration only. Its effects will be removed.
Seitenansicht 327
1 2 ... 323 324 325 326 327 328 329 330 331 332 333 ... 452 453

Kommentare zu diesen Handbüchern

Keine Kommentare