Operation
Using Probes with an Agilent 83491A
To compensate for a passive probe
1
Connect the probe to the
Input
connector on the Agilent 83491A clock recovery
module.
2
Attach the probe tip to the
CAL
hook that is located near the floppy disk drive.
3
Press the
reference receiver module’s
front-panel channel
SETUP
key.
4
Press
Calibrate
and then
Calibrate probe
.
To compensate an Agilent 54701A active probe
1
Connect the Agilent 83491A output to the electrical measurement channel
input.
2
Connect the probe to the
Input
connector on the Agilent 83491A clock recovery
module.
3
Connect the probe power cable to the
Probe Power
connector on the reference
receiver module.
4
Attach the probe tip to the
CAL
hook that is located near the floppy disk drive.
5
Press the
reference receiver module’s
front-panel channel
SETUP
key.
6
Press
Calibrate
and then
Calibrate probe
.