3. What do I need for proper cleaning?
Standard Cleaning Equipment
Before you can start your cleaning procedure you need
the following standard equipment:
• Dust and shutter caps
• Isopropyl alcohol
• Cotton swabs
• Soft tissues
• Pipe cleaner
• Compressed air
Dust and shutter caps
All Agilent Technologies lightwave instruments are de-
livered with either laser shutter caps or dust caps on the
lightwave adapter. Any cables come with covers to pro-
tect the cable ends from damage or contamination.
We suggest these protective coverings should be kept on
the equipment at all times, except when your optical dev-
ice is in use. Be careful when replacing dust caps after
use. Do not press the bottom of the cap onto the fiber too
hard, as any dust in the cap can scratch or pollute your
fiber surface.
If you need further dust caps, please contact your nearest
Agilent Technologies sales office (see Appendix, page 40).
Isopropyl alcohol
This solvent is usually available from any local pharma-
ceutical supplier or chemist’s shop.
If you use isopropyl alcohol to clean your optical device,
do not immediately dry the surface with compressed air
(except when you are cleaning very sensitive optical dev-
ices). This is because the dust and the dirt is solved and
will leave behind filmy deposits after the alcohol is evapo-
rated. You should therefore first remove the alcohol and
the dust with a soft tissue, and then use compressed air to
blow away any remaining filaments.
9
Seitenansicht 8