10
1. Zero and calibrate the power
meter and power sensor.
NOTE If you are using a power meter/sensor combination
other than the Agilent E4416A and E4413A, it may
be necessary to:
a. Enter the appropriate power sensor calibration
factors into the power meter.
b. Enable the power meter’s cal factor array.
For operating information on your particular power
meter/sensor, refer to their operating guides.
1. Connect the equipment as shown in Figure 3-5.
NOTE During the user flatness correction process, the
power meter is slaved to the signal generator via
GPIB. No other controllers are allowed on the
GPIB interface.
Configure the power meter
Figure 3-5.
User Flatness
Correction Equipment
Setup.
Connect the equipment