Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Using a Timing Analyzer and an Oscilloscope
Configure the Oscilloscope
Depending on the kind of glitch you are triggering on, the type of
probes used, and the speed of your system, your configuration could
vary from the one shown. In this example, we are triggering on a
positive glitch that is present for less that 20 ns.
Channel Setup
1. From the Channels tab in the oscilloscope window, select Setup....
2. From the Channels Setup dialog, configure channel 1 as shown below. Turn
channel 2 off.
Trigger Setup
1. From the Channels tab in the oscilloscope window, select Trigger....
2. Set the Mode field to Pattern.
3. Configure the pattern as a high pulse on channel 1.
4. Qualify the pattern to be present for less than 20 ns.
5. Select Close.