Agilent Technologies E5500 Phase Noise Measurement System
11-15
FM Discriminator Measurement Examples
FM Discriminator Measurement using Double-Sided Spur Calibration
The segment data will be displayed on the computer screen as the data is
taken until all segments have been taken over the entire range you specified
in the Measurement definition’s Type and Range.
When the
Measurement is
Complete
When the measurement is complete, refer to Chapter 15, “Evaluating Your
Measurement Results” for help in evaluating your measurement results. (If
the test system has problems completing the measurement, it will inform you
by placing a message on the computer display.
Figure 11-3 on page 11-15 shows a typical absolute measurement using FM
discrimination.
Figure 11-3 Typical Phase Noise Curve using Double-Sided Spur Calibration.