Module I/O Specifications
A9 Global Test and Demod
A9 Global Test and Demod
Use extender card 08922-60133.
Power Supplies
Inputs
The 10.7 MHz is orginated from the A16 Receiver. It is down converted to 700 kHz
± 50 kHz within the Global Test and Demod assembly. To obtain a reading either with an
oscilloscope (See Figure 4-2 on page 12-16) or spectrum analyzer, the signal needs to be
looped back. This can be done using the RF diagnostics and pausing on test #1 for
Receiver Down converters (with spectrum analyzers).
If the generator path is in doubt, use a known working reference signal into a port,
remembering to check settings on RF analyzer page for frequency and port settings.
+15 V J2(29,30) - TP 15 140 mA
-15 V J2(25,26) 80 mA
+5 V J2(23,24) 1.0 A
-5 V J2(21,22) 20 mA
Ground J1(1,33,4,17,18,20)
J3(1-4,6-14,16-20) - TP 14/16
From A16 Receiver
10.7M_IF J1(7)
Frequency: 10.7 MHz ± 50 kHz
Level: 3 dBm ± 1 dB
Harmonics: <- 40 dBc