Performance Test 4
U3401A User’s and Service Guide 65
Diode Verification Test
Configuration: Diode
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in Table 4- 5.
Provide the input shown in Table 4- 5.
3 Make a measurement and observe the result. Compare measurement
results to the appropriate test limits shown in Table 4- 5. (Be certain to
allow for appropriate source settling when using the Fluke 5520A).
Frequency Verification Test
Configuration: Frequency
1 Select the frequency function.
2 Select each range in the order shown in Table 4- 6. Provide the input
voltage and frequency indicated. Compare measurement results to the
appropriate test limits shown in Table 4- 6. (Be certain to allow for
appropriate source settling).
Table 4 - 5 Diode verification test
Function Voltage Range Error from nominal one year
Diode 1.0000 V 2.3 V ±1 mV
2.0000 V 2.3 V ±1.5 mV
Table 4 - 6 Frequency verification test
Function Voltage Input frequency Range Error from nominal
one year
Frequency 2 V 4.5 kHz 5 kHz ±0.7 Hz