6 SCPI Command Reference
106 N9340A User’s Guide
in the command. A 1 is returned only if marker
count is on and the selected number is the active
marker.
Marker Function
:CALCulate:MARKer[1]|2|3|4|5|6:FUNCtion
FCOunt|NOISe|OFF
:CALCulate:MARKer[1]|2|3|4|5|6:FUNCtion?
This command selects the marker function for the
specified marker.
FCOunt is a frequency counter function.
NOISe is a noise measurement.
OFF turns off all functions.
Marker Peak (Maximum) Search
:CALCulate:MARKer[1]|2|3|4|5|6:MAXimum
This command performs a peak search based on
the search mode settings.
Remarks: If a frequency count x value is
generated when the frequency
count state is off, then 9e15 is
return.
*RST State: Off
Key Access: [Marker] > {More} > {Freq Count}