Power Sweep Mode
Power Sweep mode is used in a power level calibration setup when the frequency is constant
(CW frequency) and the amplitude of the power source signal is swept. This mode can be used
to characterize the flatness, linearity, or gain compression of a DUT.
Table 2. Procedures to configure the N1911A/12A power meter for Power Sweep mode
Step Procedures
1 Connect the power sensor to the signal generator/signal source
2 Connect the power meter TRIG OUT to the signal source TRIG IN using a
BNC cable. The same connection is applied between the signal source TRIG
OUT and the power meter TRIG IN
3 Press [Channel] on the power meter. Set the Sensor Mode to AVG Only
4 Press [Trig/Acq] on the power meter, Acqn and select either Sing Trig or
Cont Trig
5 Press Settings>Source>Ext to set the trigger source to external
6 Press 1 of 2>Output>On to allow the TTL level high to be produced at TRIG
OUT
7 Press Slope and select + or – to set the trigger edge
8 Set the trigger buffer size by sending the SENSe:BUFFer:COUNt <buffer_
size> command to the meter through SCPI
9 Press [Channel] or [Meas] on the power meter to configure the
measurement settings such as averaging, measurement frequency, offsets,
and duty cycle
10 Send an *OPC command to power meter
11 Set the power meter to Continuous Trigger mode by sending the
INITiate:CONTinuous ON command
12 Configure and set the required power sweep range and step on the signal
source accordingly
13 Set the signal source trigger input and trigger output, then start sweeping
14 Poll the status of the power meter by sending an *ESR?. *ESR will return
1 when buffering is completed. Use FETCh? to retrieve all buffered
measurements