4
Magnitude Characteristics
Absolute amplitude accuracy (R, A, B)
–20 dBm input, 23 ±5 °C . . . . . . . . . . . . . . .< ±1.5 dB (±0.9 dB typically)
Ratio accuracy (A/R, B/R)
–20 dBm input, 23 ±5 °C, IFBW ≤ 3 kHz
100 k ≤ frequency < 1 MHz . . . . . . . . . . . . .< ±1 dB (±0.6 dB typically)
frequency ≥ 1 MHz . . . . . . . . . . . . . . . . . .< ±0.5 dB (±0.3 dB typically)
Dynamic accuracy (A/R, B/R)
Input level
(relative to full scale input level)
1
Dynamic accuracy
2
0 dB < ±0.3 dB
–10 dB to –70 dB < ±0.05 dB
–80 dB < ±0.1 dB
–90 dB < ±0.3 dB
–100 dB < ±1.0 dB
–110 dB < ±0.8 dB typically
–120 dB < ±2.5 dB typically
Residual responses
A, B inputs, frequency ≥ 3 MHz . . . . . . . . . . . . . . . . .< –95 dBm typically
R input, frequency ≥ 3 MHz . . . . . . . . . . . . . . . . . . . .< –70 dBm typically
See “EMC” under “Others” in “Common Specifications for Network
and Spectrum Measurement.”
Trace noise
A/R, B/R measurement,
–10 dBm input, IFBW = 300 Hz . . . . . . . . . . .< 0.002 dB rms typically
Stability 0.01 dB/°C typically
Dynamic accuracy (dB)
Input level (dB)
Spec
Typical
1. Full scale input level = –5 dBm
2. At 23 ±5 °C, IFBW = 10Hz, R input = –35 dBm, Reference power level = –35 dBm
Figure 1. Magnitude dynamic accuracy