Agilent Technologies 8487D Betriebsanweisung Seite 161

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 196
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 160
Error Messages
Introduction
Agilent E4418B Users Guide 4-3
Errors are retrieved in a first in first out (FIFO) order. If more than 30
errors occur, the error queue overflows and the last error in the queue is
replaced with error -350, Queue Overflow. Any time the queue
overflows the most recent error is discarded.
Once the errors are read they are removed from the error queue. This
opens a position at the end of the queue for a new error message, if one is
subsequently detected. When all errors have been read from the queue,
further error queries return +0, No errors.
To delete all the errors in the queue from the front panel press:
, , then .
To delete all the errors in the queue remotely use:
the *CLS (clear status) command.
The error queue is also cleared when the instrument power has been
switched off.
System
Inputs
More
Error List Clear Errors
HP4402.book Page 3 Wednesday, November 29, 2000 4:46 PM
Seitenansicht 160
1 2 ... 156 157 158 159 160 161 162 163 164 165 166 ... 195 196

Kommentare zu diesen Handbüchern

Keine Kommentare