21
Display Function
Display
Size/Type 8.4 inch color LCD (TFT)
Number or pixels 640 480 (VGA)
Scale type
X axis scale Linear and Log
Y axis scale Linear and Log (depends on the sweep type)
Number of traces
Data trace 2 traces (trace A and trace B)
Memory trace 2 traces (trace A and trace B)
Split display Available (trace A: upper half, trace B: lower half)
Instrument/IBASIC display selection All Instrument, Half Instrument and half IBASIC, all IBASIC, or
Instrument and IBASIC status.
Other display function Inactive trace off, Trace accumulation, Phase expansion
Data math function Data—Memory, Data/Memory
1
, Delta%
2
, Offset
1. Complex Z-Y measurement only.
2. Except for Complex Z-Y measurement.
Marker Function
Marker type and number
Main marker One for each trace (A and B).
Sub marker Seven for each trace (A and B).
∆ marker One for each trace (A and B).
Marker search
Search type Maximum, Minimum, Target, Peak, Trace bandwidth analysis
Search track Performs search by each sweep
Marker X-axis display Sweep parameter value, Sweep elapsed time, or Relaxation
time (1/2πf)
Others Marker continuous mode, ∆ marker mode, Marker coupled mode,
Marker value substitution (Marker→), Marker zooming, Marker
list, Marker statistics, Marker signal/dc bias monitor