Test Signal
Frequency
75 kHz to 30 MHz, 100 Hz solution
Frequency accuracy
±0.01%
Signal modes
Normal—Program selected voltage or current at the
measurement terminals when they are opened or
shorted, respectively
Constant—Maintains selected voltage or current at
the device under test (DUT) independent of
changes in the device’s impedance.
Signal level
The following test signal level accuracy is specified
for an ambient temperature range of 23 °C ± 5 °C
and the test cable length is 0 m.
Mode Range Setting accuracy
Voltage Normal 5 m V
rms
to 2 V
rms
±{(8 + 0.4 f
m
)% + 1 m V
rms
}
Constant
1
10 mV
rms
to 1 V
rms
±{(6 + 0.2 f
m
)% + 1 m V
rms
}
Current Normal 200 µA
rms
to 20 mA
rms
±{(8 + 1 f
m
)% + 40 µA
rms
}
Constant
1
100 µA
rms
to 20 mA
rms
±{(6 + 0.2 f
m
)% + 40 µA
rms
}
1. When the ALC function is set to ON
For the temperature range of 0 °C to 55 °C, multi-
ply the temperature induced setting error listed in
Figure 1-5 to the test signal setting accuracy. When
test cable length is 1 m or 2 m, add the following
error due to test cable length.
0.2 f
m
L [%]
where:
f
m
= Test frequency [MHz]
L = Test cable length [m]
Output impedance
The following output impedance is specified for the
test cable length of 0 m:
(
2
)
(25 + 0.5 f
m
) ± 10 +
3
f
m
%
where:
f
m
= Test frequency [MHz]
Test signal level monitor
The following test signal level monitor accuracy is
specified for an ambient temperature range of
23 °C ± 5 °C and the test cable length is 0 m.
Mode Range Monitor accuracy
Voltage 0.01 mV
rms
– 2.000 V
rms
Current 0.001 µA
rms
– 20.00 mA
rms
For the temperature range of 0 °C to 55 °C, multi-
ply the temperature induced setting error listed in
Figure 1-5. When test cable length is 1 m or 2 m,
add the following error due to test cable length.
S
mon
0.2 f
m
L [V]
where:
f
m
= Test frequency [MHz]
L = Test cable length [m]
S
mon
= Readout value of test signal level
S
set
= Setting value of test signal level
For example,
Test frequency: 1 MHz
Test signal level: 1 V
rms
Monitor readout value: 500 mV
rms
Cable length: 1 m
Ambient temperature: 25 °C
S
mon
S
set
[V ]
4 + 0.2 f
m
100 500
{}
± +