Key Parameters and Specifications
Test frequencies:
100 Hz, 120 Hz, 1 kHz, 10 kHz, 100 kHz
Option 4263B-002 adds 20kHz
AC test signal levels:
20m–1Vrms, 5mVrms steps
Basic accuracy:
0.1%
Impedance parameters:
|Z|, R, X, |Y|, G, B, C, L, D, Q,
u
Option 4263B-001 adds transformer
measurement functions: turns-ratio,
mutual-inductance and dc-resistance
Cable length settings:
0, 1, 2, 4 meters
Bias:
1.5 and 2.0 Vdc
Error correction:
Open, short, and load
Built-in system features:
GPIB and handler interfaces
Measurement time (typical):
25 ms at best conditions
Contact check time (typical):
5 ms per measurement
Display:
LCD with back-light.
Displays measurement values,
instrument states, and comparator results
Trigger mode:
Choose internal,
external, or manual trigger
Save/recall:
Store and retrieve up
to 10 measurement states
Cable:
Extend front panel cable
length from 0 to 4 meters
Contact check:
Verify reliability of test
connections
Open/short/load:
Correct for test fixture and cable
errors
Comparator:
Select values for HIGH,
IN, and LOW testing
Frequency:
Select one of five test frequencies
I&V monitor:
Monitor actual
ac voltage and
current levels
Measurement parameter:
Select the desired test parameter
Four-terminal pair:
Reduces errors due
to cable extension