Specifications
Accuracy Specifications ±(% of reading + % of range)
1
Function Range
2
Test Current 24 Hour 90 Day 1 Year Temperature Maximum
23 °C ± 1 °C 23 °C ± 5 °C 23 °C ± 5 °C Coefficient per Lead
0 °C—18 °C Resistance
28 °C—55 °C
dc Voltage 1.0000000 mV
3
0.0025 + .0020 0.0040 + .0020 0.0050 + .0020 0.0004 + .0001
10.000000 mV
3
0.0025 + .0002 0.0040 + .0002 0.0050 + .0003 0.0004 + .0001
100.00000 mV 0.0015 + .0003 0.0030 + .0004 0.0040 + .0004 0.0004 + .00006
1.0000000 V 0.0010 + .0003 0.0025 + .0004 0.0035 + .0004 0.0004 + .00004
10.000000 V 0.0002 + .0001 0.0020 + .0004 0.0030 + .0004 0.0001 + .00002
100.00000 V
4
0.0010 + .0004 0.0025 + .0005 0.0035 + .0005 0.0004 + .00005
Resistance
5
1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω
10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω
100.00000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω
1.0000000 KΩ 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω
10.000000 KΩ 100 µA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 KΩ
100.00000 KΩ 10 µA 0.0015 + .0003 0.0040 + .0004 0.0060 + .0004 0.0005 + .00002 1 KΩ
1.0000000 MΩ 5 µA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ
Low Power 1.0000000 Ω 10 mA 0.0015 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω
Resistance
5
10.000000 Ω 10 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 1 Ω
100.00000 Ω 1 mA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 10 Ω
1.0000000 KΩ 100 µA 0.0015 + .0002 0.0040 + .0002 0.0060 + .0002 0.0005 + .00001 100 Ω
10.000000 KΩ 10 µA 0.0015 + .0004 0.0040 + .0004 0.0060 + .0004 0.0005 + .00001 1 KΩ
100.00000 KΩ 5 µA 0.0015 + .0012 0.0040 + .0015 0.0060 + .0015 0.0005 + .00003 1 KΩ
1.0000000 MΩ 5 µA 0.0020 + .0003 0.0050 + .0004 0.0070 + .0004 0.0006 + .00003 1 KΩ
Voltage Limited 10.000000 Ω 1 mA 0.0020 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 1 Ω
Resistance
5, 6
100.00000 Ω 100 µA 0.0025 + .0002 0.0050 + .0002 0.0070 + .0002 0.0005 + .00002 5 Ω
Channel 1 / Channel 2 (dcV Ratio) Ratio Error in % = Channel 1 accuracy in % + Channel 2 accuracy in %
Channel 1-Channel 2 (dcV Difference) Difference Error = Channel 1 (% of reading + % of range) + Channel 2 (% of reading + % of range)
Temperature (resolution = 0.001 °C)
SPRT
7
SPRT Probe Accuracy + 0.003°C
RTD RTD Probe Accuracy + 0.05°C
Thermistor Thermistor Probe Accuracy + 0.1°C
Thermocouple
8
Thermocouple Probe Accuracy + 0.2°C
DC Voltage Noise
9
Observation Period
Range 2-Minute 2-Minute 24-Hour
RMS Peak-Peak Peak-Peak
Noise Noise Noise
1 mV 1.3 nVrms 8 nVpp 12 nVpp
10 mV 1.5 nVrms 10 nVpp 14 nVpp
100 mV 10 nVrms 65 nVpp 80 nVpp
1 V 100 nVrms 650 nVpp 800 nVpp
10 V 450 nVrms 3 µVpp 3.7 µVpp
100 V 11 µVrms 75 µVpp 90 µVpp
DC Voltage Noise vs Source Resistance
10
Source Analog Digital
Resistance Noise Filter Filter
0 Ω 1.3 nVrms Off Med
100 Ω 1.7 nVrms Off Med
1k Ω 4 nVrms Off Med
10k Ω 13 nVrms Off Med
100k Ω 41 nVrms On Med
1M Ω 90 nVrms On Slow
1 Specifications are for Channel 1 or Channel 2, after 2-hour warm-up, resolution
at 7.5 digits (100 NPLC), with FILTERS off. RESISTANCE specifications are for
4-wire Ohms or 2-wire ohms using Null. Without Null, add 0.2 Ohms additional
error in 2-wire Ohms function. For Analog Filter ON, add 0.002% of reading.
2 20% overrange on all ranges except 5% on Voltage Limited Resistance.
3 After using Math Null. If Null is not used add 100 nanoVolts.
4 Channel 1 only.
5 Channel 1 only. Resistance measurements, for NPLC <1, add 160 µΩ rms
noise.
6 Voltage limit can be set to 20 mV (default), 100 mV, or 500 mV. Measured
resistance plus Channel 1 HI and LO lead resistance is limited to 10.5 Ω on the
10 Ω range and 105 Ω on the 100 Ω range.
7 For 25 Ω SPRT with triple-point of water check within the last 4 hours.
Without the triple-point of water check, add 0.013°C for 24 hours, add 0.035°C
for
90 day, and add 0.055°C for 1 year specifications.
8 For fixed reference junction. Add 0.3ûC for external reference junction, add
2.0ûC for internal reference junction.
9 After a 2-hour warm-up, ± 1ûC, 6.5 digits (10 PLC) with Analog Filter Off and
Digital Filter Medium (50 reading average).
2-minute rms and 24-hour noise typical. For measurements using 0.02 or 0.2
NPLC, add 800 nV rms noise.
10 Typical noise behavior for Ch 1 or Ch 2, after 2 hour warm-up, 6.5 digits (10
PLC), 2 minute observation period on 1 mV range. For peak-to-peak noise, mul-
tiply rms noise by 6.