SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
Agilent E4418B/E4419B Programming Guide 9-51
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is used with the Agilent E-series power sensors to select
one of two power ranges.
If 0 is selected, the power sensors lower range is selected.
If 1 is selected, the power sensors upper range is selected.
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO to OFF.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of
the power sensors range.
1 is returned when the upper range is selected.
0 is returned when the lower range is selected.
Query Example
Error Message
This command is used with the Agilent E-series power sensors. If one is
not connected the error -241, Hardware missing occurs.
POW:AC:RANG 0 This command sets the power
sensor to its lower range.
POW:AC:RANG? This command queries the current
setting of the power sensor range.
SENS
1 :ACPOW
SENS2
:
:RANG
Space
?
numeric_value
prog.book Page 51 Thursday, June 7, 2001 2:55 PM
Seitenansicht 352