Agilent Technologies 1000A Wartungshandbuch

Stöbern Sie online oder laden Sie Wartungshandbuch nach Multimeter Agilent Technologies 1000A herunter. Agilent Technologies 1000A Technical data Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Oscilloscopes

s1 Agilent 1000A Series OscilloscopesService Guide

Seite 2

10 Agilent 1000A Series Oscilloscopes Service Guide Tables

Seite 3

s111Agilent 1000A Series OscilloscopesService Guide 1ServiceCleaning the Oscilloscope 12Testing Performance 13Performance Test Record 35Returning the

Seite 4 - In This Book

1Service12 Agilent 1000A Series Oscilloscopes Service GuideCleaning the OscilloscopeIf the instrument requires cleaning:1 Remove power from the instr

Seite 5 - Contents

Service 1Agilent 1000A Series Oscilloscopes Service Guide 13Testing PerformanceThis section documents performance test procedures. Performance verifi

Seite 6

1Service14 Agilent 1000A Series Oscilloscopes Service GuideBefore Testing PerformanceTable 2 Equipment Required to Perform All TestsDescription Requi

Seite 7

Service 1Agilent 1000A Series Oscilloscopes Service Guide 15Calibration1 Push the [Utility] key on the front panel.2 In the Utility menu, press Self-

Seite 8

1Service16 Agilent 1000A Series Oscilloscopes Service GuideTo test DC gain accuracy1 Disconnect all cables from the oscilloscope channel inputs.2 Pre

Seite 9

Service 1Agilent 1000A Series Oscilloscopes Service Guide 176 Set the channel 1 probe attenuation to 1X.7 Set the channel 1 vertical sensitivity valu

Seite 10

1Service18 Agilent 1000A Series Oscilloscopes Service Guide10 Press [Measure].11 Press the Voltage softkey.12 Turn the entry knob to select the Va

Seite 11

Service 1Agilent 1000A Series Oscilloscopes Service Guide 1913 For each of channel 1’s vertical sensitivities in the DC Gain Test section of the “Per

Seite 12 - Cleaning the Oscilloscope

Agilent 1000A Series Oscilloscopes Service GuideNotices© Agilent Technologies, Inc. 2009, 2013No part of this manual may be reproduced in any form or

Seite 13 - Testing Performance

1Service20 Agilent 1000A Series Oscilloscopes Service Guide17 Press the [Default Setup] front panel key.18 Set the channel’s probe attenuation to 1X.

Seite 14 - Before Testing Performance

Service 1Agilent 1000A Series Oscilloscopes Service Guide 21To test analog bandwidth, maximum frequencyCAUTIONEnsure that the input voltage to the os

Seite 15 - Calibration

1Service22 Agilent 1000A Series Oscilloscopes Service GuideTable 6 Equipment Required for Analog Bandwidth TestDescription Critical Specifications Re

Seite 16 - To test DC gain accuracy

Service 1Agilent 1000A Series Oscilloscopes Service Guide 231 Connect the equipment as shown in Figure 4.2 Preset and calibrate the power meter accor

Seite 17

1Service24 Agilent 1000A Series Oscilloscopes Service Guide8 Set the horizontal scale to 500 ns/div.Figure 5 Channel 1 Vertical Scale SettingFigure 6

Seite 18

Service 1Agilent 1000A Series Oscilloscopes Service Guide 259 Press [Acquire].10 Press the Acquisition softkey until “Average” appears.11 Press Ave

Seite 19

1Service26 Agilent 1000A Series Oscilloscopes Service Guide16 Using the Vpp reading, calculate the Vrms value using the following expression and reco

Seite 20

Service 1Agilent 1000A Series Oscilloscopes Service Guide 2718 Calculate the reference gain as follows:Record this value in the Calculated Gain @ 1 M

Seite 21 - Service 1

1Service28 Agilent 1000A Series Oscilloscopes Service GuideFor example, if Pmeas = 3.65 mW:23 Calculate the gain at the maximum frequency using the e

Seite 22 - 1Service

Service 1Agilent 1000A Series Oscilloscopes Service Guide 29To test time scale accuracyIn this test you measure the absolute error of the timebase os

Seite 23

Agilent 1000A Series Oscilloscopes Service Guide 3 Agilent 1000A Series Oscilloscopes—At a GlanceThe Agilent 1000A Series oscilloscopes are low-cost p

Seite 24

1Service30 Agilent 1000A Series Oscilloscopes Service Guide1 Set up the signal generator:a Set the output to 10 MHz, approximately 1 Vpp sine wave.2

Seite 25

Service 1Agilent 1000A Series Oscilloscopes Service Guide 313 Set up the oscilloscope:a Press [Auto-Scale].b Set the oscilloscope Channel 1 probe at

Seite 26 - Vpp reading

1Service32 Agilent 1000A Series Oscilloscopes Service GuideTo test trigger sensitivityThis test verifies the trigger sensitivity. In this test, you a

Seite 27

Service 1Agilent 1000A Series Oscilloscopes Service Guide 331 Connect the equipment (see Figure 10).a Connect the signal generator output to the osci

Seite 28

1Service34 Agilent 1000A Series Oscilloscopes Service GuideThe trigger is stable when the displayed waveform is stable. If the trigger is not stable,

Seite 29 - CN XX XX XXXX

Service 1Agilent 1000A Series Oscilloscopes Service Guide 35Performance Test RecordTable 12 Performance Test InformationSerial No. __________________

Seite 30

1Service36 Agilent 1000A Series Oscilloscopes Service GuideChannel 22mV/div ±6mV +0.96 to +1.045 mV/div ±15 mV +0.96 to +1.0410 mV/div ±30 mV +0.97 t

Seite 31

Service 1Agilent 1000A Series Oscilloscopes Service Guide 37Channel 42mV/div ±6mV +0.96 to +1.045 mV/div ±15 mV +0.96 to +1.0410 mV/div ±30 mV +0.97

Seite 32 - To test trigger sensitivity

1Service38 Agilent 1000A Series Oscilloscopes Service GuideTable 15 Time Scale AccuracyLimits Measured time scale error (ppm) Pass/Fail±50 ppm from 0

Seite 33 - ≥ 5mV/div

Service 1Agilent 1000A Series Oscilloscopes Service Guide 39Returning the Oscilloscope to Agilent for ServiceBefore shipping the oscilloscope to Agil

Seite 34 - < 5mV/div

4 Agilent 1000A Series Oscilloscopes Service Guide In This BookThis guide contains service information for the Agilent 1000A Series oscilloscopes.1Ser

Seite 35 - Performance Test Record

1Service40 Agilent 1000A Series Oscilloscopes Service GuideContacting AgilentInformation on contacting Agilent Technologies can be found at www.agile

Seite 36

s141Agilent 1000A Series OscilloscopesService Guide ASafety NoticesWarnings 41Safety Symbols 42This apparatus has been designed and tested in a

Seite 37

ASafety Notices 42 Agilent 1000A Series Oscilloscopes Service Guide• Whenever it is likely that the ground protection is impaired, you must make the

Seite 38 - Table 16 Trigger Sensitivity

Agilent 1000A Series Oscilloscopes Service Guide 43 IndexAanalog bandwidth - maximum frequency check, 21at a glance, 3Bbuilt-in help, 3Ccalibration, 1

Seite 39

44 Agilent 1000A Series Oscilloscopes Service GuideIndex

Seite 41 - Safety Notices

s1© Agilent Technologies, Inc. 2009, 2013Available in electronic format onlyThird Edition, April 2013*54130-97015*54130-97015www.agilent.com

Seite 42 - Safety Symbols

Agilent 1000A Series Oscilloscopes Service Guide 5 ContentsAgilent 1000A Series Oscilloscopes—At a Glance 3In This Book 4Figures 7Tab les 91Service11C

Seite 43

6 Agilent 1000A Series Oscilloscopes Service Guide Contents

Seite 44

Agilent 1000A Series Oscilloscopes Service Guide 7 FiguresFigure 1. Averages Menu Item 17Figure 2. Connecting Equipment for DC Gain Accuracy Test 18Fi

Seite 45

8 Agilent 1000A Series Oscilloscopes Service Guide Figures

Seite 46 - *54130-97015*

Agilent 1000A Series Oscilloscopes Service Guide 9 Tab lesTable 1. Agilent 1000A Series Oscilloscope Models 3Table 2. Equipment Required to Perform Al

Kommentare zu diesen Handbüchern

Keine Kommentare