50 User’s Guide
U3025AE10 Operational Check
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in Table 12 on
page 43. This can be done using Limit Lines (press Trace > Limit Test) or Marker. The
trace should be above the limit line and PASS will be displayed on the screen. Refer to
Figure 48.
11.Repeat step 9 and step 10 for Cal Sets “999.1” thru “999.12” (12-Port).
Figure 48 E8364B/C Reflection Tracking Trace (Ports 1-12)