25
N5230A/C Corrected System Performance with 2.4mm Connectors
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
85056A Calibration Kit:
⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯
Table 9. 85056A Calibration Kit
N5230A/C- Option 420 or 520 (Standard Test Set and Standard Power Range)
Applies to the N5230A/C Option420 or 520 analyzers, 85056A (2.4mm) Electronic Calibration Module, 85133F flexible test port cable set,
and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
40 GHz
40 to
50 GHz
Directivity 42 42 38 36
Source Match 41 38 33 31
Load Match 42 42 37 35
Reflection Tracking
±0.001
+0.02/°C
±0.008
+0.02/°C
±0.020
+0.02/°C
±0.027
+0.03/°C
Transmission Tracking
±0.019
+0.02/°C
±0.060
+0.02/°C
±0.129
+0.02/°C
±0.211
+0.03/°C