Performance Verification 5
Average Gain of Zoomed-In Channel
N2820/1A User’s Guide 75
positive gain:
7 Set the DC power supply to –20 mV.
NOTE Although this is the recommended setting, you can use any negative
voltage less than –35 mV.
8 Connect the multimeter on the DC power supply as shown
in Figure 43 on page 73.
9 Record the value of –V
in
measured on the multimeter:
–V
in
:
10 Connect the multimeter’s leads to the banana plug on the
output of the N1022A/B as shown in Figure 44 on page 74.
11 Record the value of –V
out
measured on the multimeter:
–V
out
:
12 Calculate the negative gain of the zoomed- in (primary)
channel and record the result in the following line:
negative gain:
13 Calculate the average of the two gains:
positive gain
+V
out
+V
in
------------
=
negative gain
–V
out
–V
in
-----------
=