Agilent Technologies HP 8566B Spezifikationen Seite 56

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 405
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 55
9. Calibrator Amplitude Accuracy Test
9.
Calibrator
Amplitude
Accuracy
Test
Related
Adjustment
CAL OUTPUT Adjustment
Specification
-10
dBm
f0.3
dB;
100 MHz
Description
The output level of the calibrator signal is measured with a power
meter.
SPECTRUM ANALYZER
POWER METER
POWER
SENSOR
Figure 2-13. Calibrator Amplitude Accuracy
Test
Setup
Equipment
Power Meter
...............................................
..HP436
A
Power Sensor
.............................................
..HP8481
A
Adapter, Type N (f) to BNC (m)
...........................
.1250-1477
Procedure
1. Connect equipment as shown in Figure 2-13.
2. Measure output level of the CAL OUTPUT signal. The value should
be
-
10.0
dBm
f0.3
dB.
dBm
Performance Tests 2-31
Seitenansicht 55
1 2 ... 51 52 53 54 55 56 57 58 59 60 61 ... 404 405

Kommentare zu diesen Handbüchern

Keine Kommentare