3. Check Channels Leakage
Send CLOS(@nn00) to the FET Multiplexer, where nn is the
channel number (typically 01)
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to Chassis, Channels
Send OPEN (@nn00) to the FET Multiplexer
4. Check Tree Leakage Current
Send SCAN:PORT ABUS to the FET Multiplexer
Send CLOS(@nn00) to the FET Multiplexer to close channel
00 and Tree A, where nn is the card number (typically 01)
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to Chassis, Tree A
Send OPEN (@nn00) to the FET Multiplexer
Agilent E1300B/E1301B
Figure 2-8. Positive HI to Chassis Leakage Connections
Verification Tests 2-15