S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec
Equivalent Front-Panel Key Commands
INCR
×
10
The INCR×10 Data Function increases the increment setting by a factor of 10
(new increment setting = current increment setting
× 10).
NOTE: Integer-only fields (such as Intensity and Print Adrs) have a fixed
increment of 1, and cannot be changed.
Syntax
:INCRement:MULTiply
Example
OUTPUT 714;"RFG:FREQ:INCR:MULT"
If the RF Generators frequency increment is 1 MHz, this command
increases increment value from 1 MHz to 10 MHz.
INCR
÷
10
The INCR÷10 Data Function reduces the increment setting by a factor of 10 (new
increment setting = current increment setting
÷ 10).
NOTE: Integer-only fields (such as Intensity and Print Adrs) have a fixed
increment of 1, and cannot be changed.
Syntax
:INCRement:DIVide
Example
OUTPUT 714;"RFG:FREQ:INCR:DIV"
If the RF Generators frequency increment is 10 MHz, this command
reduces the increment value from 10 MHz to 1 MHz.